Norovirus (GⅠ) RT-PCR Detection Kit
Product Name
Norovirus (GⅠ) RT-PCR Detection Kit (Lyophilized)
Size
48tests/kit, 50tests/kit
Intended Use
It is suitable for the detection of Norovirus (GⅠ) in shellfish, raw vegetables and fruits, water, feces, vomit and other specimens. Nucleic acid extraction should be carried out by nucleic acid extraction kit or direct pyrolysis method according to different specimen types.The kit is an ALL-READY PCR SYSTEM(Lyophilized), which contains the DNA amplification enzyme, reverse transcriptase,reaction buffer, specific primers and probes required for fluorescent PCR detection.
Product Contents
Components | Package | specification | Ingredient |
Norovirus (GⅠ) PCR Mix | 1 ×bottle (Lyophilized powder) | 50Test | dNTPs, MgCl2, Primers, Probes,Reverse Transcriptase,Taq DNA polymerase |
6×0.2ml 8 well-strip tube(Lyophilized) | 48Test | ||
Positive Control | 1*0.2ml tube (lyophilized) | 10Tests |
Plasmid containing Norovirus (GⅠ) specific fragments |
Dissolving solution | 1.5 ml Cryotube | 500uL | / |
Negative Control | 1.5 ml Cryotube | 200uL | 0.9%NaCl |
Storage & Shelf Life
(1)The kit can be transported at room temperature.
(2)The shelf life is 18 months at -20℃ and 12 months at 2℃~30℃.
(3)See the label on kit for the production date and expiration date.
(4)The lyophilized powder version reagent should be stored at -20℃ after dissolution and the repeated freeze -thaw should be less than 4 times.
Instruments
GENECHECKER UF-150, UF-300 real-time fluorescence PCR instrument.
Operation Diagram
a) Bottle version:
b) 8 well-strip tube version:
Pcr Amplification
Recommended Setting
Step | Cycle | Temperature (℃) | Time | Fluorescence channel |
1 | 1 | 50 | 8min | |
2 | 1 | 95 | 2min | |
3 | 40 | 95 | 5s | |
60 | 10s | Collect FAM fluorescence |
*Note: The signals of FAM fluorescence channels will be collected at 60℃.
Interpreting Test Results
Channel |
Interpretation of results |
FAM Channel |
|
Ct≤35 |
Norovirus (GⅠ) Positive |
Undet |
Norovirus (GⅠ) Negative |
35<Ct≤40 |
Suspicious resut, retest* |
*If the retest result of FAM channel has a Ct value ≤40 and shows typical “S” shape amplification curve, the result is interpreted as positive, otherwise it is negative.