Staphylococcus aureus PCR Detection Kit

Short Description:

Staphylococcus aureus belongs to the genus Staphylococcus and is a gram-positive bacteria. It is a common food-borne pathogenic microorganism that can produce enterotoxins and cause food poisoning.


Product Detail

Product Tags

Product Name

Staphylococcus aureus PCR Detection Kit (lyophilized)

Size

48tests/kit, 50tests/kit

Intended Use

Staphylococcus aureus belongs to the genus Staphylococcus and is a gram-positive bacteria. It is a common food-borne pathogenic microorganism that can produce enterotoxins and cause food poisoning. This kit uses the principle of real-time fluorescent PCR and is suitable for the qualitative detection of Staphylococcus aureus in food, water samples, feces, vomit, enrichment liquid and other samples.

Product Contents

Components Package specification Ingredient
PCR Mix 1 ×bottle                  (Lyophilized powder)  50Test dNTPs, MgCl2, Primers, Probes,Taq DNA polymerase
6×0.2ml 8 well-strip tube(Lyophilized) 48Test
Positive Control 1*0.2ml tube (lyophilized)  10Tests

Plasmid or Pseudovirus containing specific fragments

Dissolving solution 1.5 ml Cryotube 500uL /
Negative Control 1.5 ml Cryotube 100uL 0.9%NaCl

Storage & Shelf Life

(1)The kit can be transported at room temperature.

(2)The shelf life is 18 months at -20℃ and 12 months at 2℃~30℃.

(3)See the label on kit for the production date and expiration date.

(4)The lyophilized powder version reagent should be stored at -20℃ after dissolution and the repeated freeze -thaw should be less than 4 times.

Instruments

GENECHECKER UF-150, UF-300 real-time fluorescence PCR instrument.

Operation Diagram

a) Bottle version:

1

b)  8 well-strip tube version:

2

Pcr Amplification

Recommended Setting

Step

Cycle

Temperature (℃)

Time

Fluorescence channel

1

1

95

2min

/

2

40

95

6s

/

60

12s

Collect FAM fluorescence

Interpreting Test Results

Channel

Interpretation of results

FAM Channel

Ct≤35

Staphylococcus aureus Positive

Undet

Staphylococcus aureus Negative

35<Ct40

Suspicious resut, retest*

*If the retest result of FAM channel has a Ct value ≤40 and shows typical S shape amplification curve, the result is interpreted as positive, otherwise it is negative.