Vibrio parahaemolyticus PCR Detection Kit
Product Name
Vibrio Parahaemolyticus PCR Detection Kit (Lyophilized)
Size
48tests/kit, 50tests/kit
Intended Use
Vibrio Parahemolyticus (also known as Halophile Vibrio Parahemolyticus) is a Gram-negative polymorphic bacillus or Vibrio Parahemolyticus.with acute onset, abdominal pain, vomiting, diarrhea and watery stool as the main clinical symptoms.This kit uses the principle of real-time fluorescent PCR and is suitable for the qualitative detection of Vibrio Parahaemolyticus in food, water samples, feces, vomit, and enrichment fluid.The kit is an ALL-READY PCR SYSTEM(Lyophilized), which contains the DNA amplification enzyme, reaction buffer, specific primers and probes required for fluorescent RT-PCR detection.
Product Contents
Components | Package | specification | Ingredient |
PCR Mix | 1 ×bottle (Lyophilized powder) | 50Test | dNTPs, MgCl2, Primers, Probes,Taq DNA polymerase |
6×0.2ml 8 well-strip tube(Lyophilized) | 48Test | ||
Positive Control | 1*0.2ml tube (lyophilized) | 10Tests |
Plasmid or Pseudovirus containing specific fragments |
Dissolving solution | 1.5 ml Cryotube | 500uL | / |
Negative Control | 1.5 ml Cryotube | 100uL | 0.9%NaCl |
Storage & Shelf Life
(1)The kit can be transported at room temperature.
(2)The shelf life is 18 months at -20℃ and 12 months at 2℃~30℃.
(3)See the label on kit for the production date and expiration date.
(4)The lyophilized powder version reagent should be stored at -20℃ after dissolution and the repeated freeze -thaw should be less than 4 times.
Instruments
GENECHECKER UF-150, UF-300 real-time fluorescence PCR instrument.
Operation Diagram
a) Bottle version:
b) 8 well-strip tube version:
Pcr Amplification
Recommended Setting
Step |
Cycle |
Temperature (℃) |
Time |
Fluorescence channel |
1 |
1 |
95 |
2min |
/ |
2 |
40 |
95 |
6s |
/ |
60 |
12s |
Collect FAM fluorescence |
Interpreting Test Results
Channel |
Interpretation of results |
FAM Channel |
|
Ct≤35 |
Vibrio Parahaemolyticus Positive |
Undet |
Vibrio Parahaemolyticus Negative |
35<Ct≤40 |
Suspicious resut, retest* |
*If the retest result of FAM channel has a Ct value ≤40 and shows typical “S” shape amplification curve, the result is interpreted as positive, otherwise it is negative.